External quality assessment (EQA) with commutable samples is essential for assessing the quality of assays performed by laboratories, particularly when the emphasis is on their standardization status and interchangeability of results.
In this study a panel of 20 fresh-frozen single-donation serum samples were used to assess assays for the measurement of creatinine, glucose, phosphate, uric acid, total cholesterol, HDL cholesterol, LDL cholesterol, and triglycerides. The commercial random access platforms included: Abbott Architect, Beckman Coulter AU, Ortho Vitros, Roche Cobas, Siemens Advia, and Thermo Scientific Konelab.
Most assays showed excellent peer performance attributes, except for HDL- and LDL cholesterol. Cases in which individual assays had biases exceeding the used limits were the Siemens Advia creatinine (−4.2%), Ortho Vitros phosphate (8.9%), Beckman Coulter AU triglycerides (5.4%), and Thermo Scientific Konelab uric acid (6.4%), which lead to considerable interassay discrepancies. Additionally, large laboratory effects were observed that caused interlaboratory differences of >30%.